CONTRIBUTED ARTICLE
5. Study of Sample Heating Effects Arising
During Laser Raman Spectroscopy
Yvonne D. West
Department of Chemistry, University of Southampton,
Highfield, Southampton, SO17 1PF.
Keywords
Raman, Sample heating, Sampling techniques
Abstract
Relatively low laser
powers (125mW) can significantly raise the temperature of a heat sensitive sample.
Furthermore, this can result in large temperature gradients across the sampled area. These
effects can be reduced by using liquid paraffin mulls and pressed disks as commonly
employed in IR studies.
|
. . . . . . . . . . . . . . . . . . . . . . . .
[ijvs/volume1/edition1/section2.html#page20] . . . . . . . . . . . . . . . . . . . . . . .
.
Introduction

Sample heating is a well known problem associated with laser Raman spectroscopy [1] because the sample absorbs some of the incident radiation. In
general, the temperature rise depends on quantities such as the thermal conductivity and
colour of the sample as well as the laser power and sampling time. To minimise the effect
requires an understanding of the amount of sample heating associated with particular
experimental conditions and a knowledge of sampling methods available for dissipating the
heat. By reducing heating effects, Raman spectra can be recorded for a range of samples
which may otherwise undergo a phase transition, decompose or emit radiation as a result of
being coloured.
In theory, it is possible to determine the temperature at the point of sampling directly
from the Raman spectrum using the ratio of the antiStokes and Stokes intensities for a
particular vibration and assuming a Boltzmann distribution for the levels involved (v=0
and v=1) [2]. The required formula is:
where v0 is the excitation frequency (cm-1), v is the vibrational frequency (cm-1),
k is Boltzmann's constant, T is the temperature and is the
vibrational energy gap (J/molecule). However, it is often not convenient, can prove very
difficult or may even be impossible to measure this ratio.
Therefore until recently, it has proven difficult to quantify the amount of heating,
although a previous study on PTFE estimated a 1°C temperature rise per 100mW laser power
[3]; this is considered to be relatively imprecise.
Sodium Molybdate undergoes a fairly sharp, non-reversible crystal transition at about
112°C (differential scanning calorimetry). The transition, which can be monitored by
changes in the Raman spectrum [4], is observed at apparently
much lower bulk sample temperatures, suggesting sample heating by the laser. A more
detailed study of the transition is described which details the effects of laser heating
as a function of apparent (bulk) sample temperature, laser power and sampling time.
It has been possible to record the spectra of some temperature sensitive samples by using
cells specially designed to reduce heating, e.g. rotating the sample. An alternative
approach is to surround the sample by a good thermal conductor [5].
A selection of highly coloured solid samples, most of which have not previously been
studied by n.i.r FT Raman spectroscopy, have been used to investigate such dissipative
sampling techniques. In particular, the studies have focused on nujol (liquid paraffin)
mulls between KBr, quartz and Pyrex flats and pressed disks composed with KBr and KCl.
. . . . . . . . . . . . . . . . . . . . . . . .
[ijvs/volume1/edition1/section2.html#page21] . . . . . . . . . . . . . . . . . . . . . . .
.
Experimental

All Raman spectra were recorded using a Perkin-Elmer 1760 FT Raman spectrometer equipped
with a Spectron c.w. Nd3+:YAG laser with TEM00 output at 1064nm.
Experiments to study the amount of laser heating used samples of standard anhydrous sodium
molybdate in a Ventacon hot cell, described elsewhere [1]. Fresh
samples were used for each independent experiment and special care was taken to ensure
temperature equilibration before recording spectra at 1cm-1
resolution. The experiments studied the effect of changing one variable (laser power, bulk
sample temperature or time of laser illumination) by recording a series of Raman spectra
and examining the differences between them. The selected experimental conditions were: (i)
Varying the laser power to cover the range 50 to 875mW, co-adding 5 scans and setting a
sample temperature of 70°C. (ii) Varying the number of scans from 10 to 70 in increments
of 10 scans, using a laser power of 125mW and setting a sample temperature of 70°C. (iii)
Varying the sample temperature in the range 20 to 80°C, co-adding 5 scans and using a
laser power of 250mW.
The experiments to investigate methods of sample cooling by thermal transfer in the
surrounding medium used highly coloured powdered samples of KMnO4
(purple), graphite (black), NiSO4.xH2O
(green), CuSO4.5H2O
(blue), Cu(dpta)(NO3)2
(blue), Cu(dien)(ClO4)2
(blue), Cu(dien)(NO3)2
(blue), Cu(dien)Br2 (blue), Cu(dpta)(NCS)(ClO4) (blue) and FeCl3 (very
dark yellow; appears black) which are all very heat sensitive [dpta=diphenyltetraamine;
dien=diehtylenetriamine]. The Raman spectrum of each sample was recorded at room
temperature as the pure sample and using a selection of the following preparations (i)
nujol mull between KBr flats, (ii) nujol mull between Pyrex flats, (iii) nujol mull
between quartz flats, (iv) pressed disk made from KBr and (v) pressed disk made from KCl.
The mull and disk preparation followed the standard procedures used for IR although the
disks were, in general, thicker and contained a higher concentration of sample than would
be typical for IR use. In each case, the resolution was 4cm-1
(except KMnO4 where 1cm-1
resolution was used), but the laser power and number of scans were optimised to give the
best spectrum. Furthermore, the effectiveness of a nujol mull in dissipating heat was
studied as a function of sample thickness between the flats, mull stodginess and particle
size for NiSO4.xH2O.
The experiment to study spectral variations with sample thickness (0.25 to 2.5mm) used a
quartz/nujol mull, 250mW laser power and 4cm-1
resolution for 50, 100 and 200 scans. Different proportions of sample to nujol in the mull
(stodginess factor; mean consistency similar to toothpaste) were also tried. A single
stock mull was used for a related series of spectra. Additional studies on the effect of
particle size used a sample (made from a stock solid/nujol mixture) defined as coarse (as
provided), medium (partially ground) or fine (well ground). The spectra were recorded as a
KBr/nujol mull, using a laser power of 200mW, 4cm-1
resolution and 500 scans.
. . . . . . . . . . . . . . . . . . . . . . . .
[ijvs/volume1/edition1/section2.html#page22] . . . . . . . . . . . . . . . . . . . . . . .
.
Results and Discussion

The results are presented in two sections comprising a discussion on measuring the amount
of laser heating followed by an evaluation of the sampling methods for dissipating the
heat. If a sample undergoes heating in the presence of incident laser radiation there are
a number of possible spectral consequences. The most common effect is the appearance of a
broad background feature which has a maximum intensity at the Raman detector cutoff for a
near infrared system ( 3600cm-1) and which falls across the
spectrum towards the low frequency Raman shift (laser line frequency); the affected region
depends on the size of the background feature which is in turn directly related to the
amount of laser heating.
For some samples, this feature remains constant for the duration of the scan, in which
case a distorted spectrum is obtained (NiSO4.xH2O) which may or may not be acceptable. Often, the background
feature continues to gain intensity during the scan as the sample continues to be heated
with the eventual result that it completely dominates the spectrum (graphite). Sometimes
the broad background feature is not observed but laser heating of the sample is still
occurring. The eventual result may be sample decomposition and hence a loss of the
spectrum of interest (KMnO4).
Amount of Laser Heating

The crystal transition in Na2MoO4
can easily be observed as changes in the Raman spectrum. Particularly significant is the
difference in the strongest band assigned as the symmetric stretch, v1, according to the tetrahedral symmetry of the free
molybdate ion. In the low temperature crystal form there are two peaks at 891 and 896cm-1 but in the high temperature form there is only one at 890cm-1.
The laser produces a significant sample heating effect and this is illustrated in figure 1
which shows spectra recorded at 250mW laser power, 1cm-1
resolution and 5 scans for varying sample temperatures. The series clearly demonstrates
the gradual change from the pure low temperature form at 20°C to the pure high
temperature form at 80°C. At intermediate temperatures, only part of the sample region
has undergone the crystal transition; there is a qualitative relationship between the
sample temperature and the proportion of sample which is in the high temperature form.
. . . . . . . . . . . . . . . . . . . . . . . .
[ijvs/volume1/edition1/section2.html#page23] . . . . . . . . . . . . . . . . . . . . . . .
.
Figure 1: Spectra of Na2MoO4
recorded using a laser power of 250mW, 1cm-1 resolution,
five scans and sample temperatures of: (a) 20°C, (b) 40°C, (c) 55°C, (d) 70°C and (e)
80°C.
[High-Resolution Spectrum]
. . . . . . . . . . . . . . . . . . .
. . . . . [ijvs/volume1/edition1/section2.html#page24] . . . . . . . . . . . . . . . . . .
. . . . . .
Figure 2 shows a series of spectra recorded using a sample temperature of 70°C, 1cm-1 resolution and 5 scans for varying laser powers. As the
laser power is increased there is a greater sample heating effect in a given time. It is
significant that low laser powers (under 200mW) can be responsible for substantial heating
effects of at least 40°C over parts of the sampled area. Furthermore, there appears to be
a large temperature gradient over the sample surface. The temperature gradient should be a
function of the laser beam diameter and coherence as well as the spectrometer resolution.
Increasing the spectrometer resolution would decrease the area of sample viewed and
therefore reduce the apparent temperature inhomogeneity. The amount of sample heating due
to the laser beam gradually diminishes with increasing distance from the central impact
point.
Figure 2: Spectra of Na2MoO4
recorded for a sample temperature of 70°C, 1cm-1
resolution, five scans and laser powers of: (a) 125mW, (b) 375mW, (c) 625mW and (d) 875mW.
[High-Resolution Spectrum]
. . . . . . . . . . . . . . . . . . .
. . . . . [ijvs/volume1/edition1/section2.html#page25] . . . . . . . . . . . . . . . . . .
. . . . . .
A further study was undertaken to monitor the changes in the spectrum as a function of
time of illumination by the laser beam. The effects are shown in figure 3 using a laser
power of 125mW, sample temperature of 70°C, 1cm-1
resolution and a varying number ofscans (illumination time). As the time of illumination
increases, the sample continues to heat to a maximum temperature with a rate dependent on
the position relative to the centre of the laser beam. Eventually, it can be observed that
all the sampled region is in the high temperature crystal form; the time to achieve this
varies from 10 minutes at 125mW to less than 2 minutes at 725mW [1].
Figure 3: Spectra of Na2MoO4
recorded at a sample temperature of 70°C, 1cm-1
resolution and using a laser power of 125mW for: (a) 10 scans, (b) 20 scans, (c) 40 scans
and (d) 70 scans.
[High-Resolution Spectrum]
. . . . . . . . . . . . . . . . . . .
. . . . . [ijvs/volume1/edition1/section2.html#page26] . . . . . . . . . . . . . . . . . .
. . . . . .
Sampling Techniques to Minimise Laser Heating Effects

The samples chosen for the study are all coloured powders which are difficult to study by
Raman spectroscopy because they tend to undergo significant heating in the laser beam and
also often absorb the Raman scatter. The techniques used are well known, readily available
and involve simple sample preparation. The spectra recorded are summarised in table 1 and
a selection, chosen to illustrate the success with different coloured solid samples in a
nujol mull, are shown in figure 4.
| Sample |
Solid |
KBr flats/nujol mull |
KBr pressed disk |
KCl pressed disk |
Pyrex flats/nujol mull |
Quartz flats/nujol mull |
| KMnO4 |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
|
![[tick]](tick.gif) |
![[tick]](tick.gif) |
| Graphite |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
|
|
| NiSO4.xH2O |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
|
![[tick]](tick.gif) |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
| CuSO4.5H2O |
![[cross]](cross.gif) |
![[tick]](tick.gif) |
|
![[cross]](cross.gif) |
|
|
| FeCl3 |
![[cross]](cross.gif) |
|
|
|
|
![[tick]](tick.gif) |
| Cu(dpta)(NO3)2 |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
|
![[tick]](tick.gif) |
|
| Cu(dien)(ClO4)2 |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
![[tick]](tick.gif) |
|
|
|
| Cu(dien)(NO3)2 |
![[cross]](cross.gif) |
![[tick]](tick.gif) |
|
|
|
|
| Cu(dien)Br2 |
![[cross]](cross.gif) |
![[tick]](tick.gif) |
|
|
|
|
| Cu(dpta)(NCS)(ClO4) |
![[cross]](cross.gif) |
![[tick]](tick.gif) |
|
|
|
|
Table 1: Sampling techniques used to record the Raman spectra of some coloured
powders.
No entry indicates no attempt was made to record the spectrum. A " " indicates a spectrum was
successfully recorded and a " " indicates attempts to record the spectrum were
unsuccessful.
. . . . . . . . . . . . . . . . . . . . . . . .
[ijvs/volume1/edition1/section2.html#page27] . . . . . . . . . . . . . . . . . . . . . . .
.
Figure 4: Spectra of different coloured powder samples recorded as nujol mulls
between KBr disks:
(a) KMnO4 (purple), 800mW laser power, 1cm-1 resolution, 10 scans;
(b) NiSO4.xH2O
(green), 200mW laser power, 4cm-1 resolution, 500 scans;
(c) Cu(dien)(NO3)2
(blue), 250mW laser power,4cm-1 resolution, 200 scans;
(d) FeCl3 (very dark yellow), 45mW laser power, 4cm-1 resolution, 500 scans.
[High-Resolution Spectrum]
. . . . . . . . . . . . . . . . . . .
. . . . . [ijvs/volume1/edition1/section2.html#page28] . . . . . . . . . . . . . . . . . .
. . . . . .
Figures 5 & 6 show the spectra recorded under the conditions indicated in table 1 for
NiSO4.xH2O and
graphite, respectively. These are typical examples. For every sample, all the techniques
tried allowed significantly better spectra to be recorded than could be obtained directly
from the pure solid. Compared with the pure solid spectra, it can be seen in figures 5
& 6 that the signal intensity increases and the noise level decreases giving greatly
improved signal-to-noise ratios, as would be expected when it is possible to have more
flexibility and therefore to optimise experimental recording conditions.
Figure 5: Spectra of NiSO4.xH2O:
(a) solid, 75mW laser power, 4cm-1 resolution, 250
scans;
(b) Pyrex flats/nujol mull, 250mW laser power, 4cm-1
resolution, 400 scans;
(c) KBr flats/nujol mull, 200mW laser power, 4cm-1
resolution, 500 scans;
(d) KCl disk, 375mW laser power, 4cm-1 resolution, 500
scans.
[High-Resolution Spectrum]
. . . . . . . . . . . . . . . . . .
. . . . . . [ijvs/volume1/edition1/section2.html#page29] . . . . . . . . . . . . . . . . .
. . . . . . .
Figure 6: Spectra of graphite:
(a) solid, 50mW laser power, 4cm-1 resolution, 500
scans;
(b) KBr flats/nujol mull, 125mW laser power, 4cm-1
resolution, 1750 scans;
(c) KBr disk, 150mW laser power, 4cm-1 resolution, 1500
scans;
(d) KCl disk, 300mW laser power, 4cm-1 resolution, 250
scans.
[High-Resolution Spectrum]
. . . . . . . . . . . . . . . . . . .
. . . . . [ijvs/volume1/edition1/section2.html#page30] . . . . . . . . . . . . . . . . . .
. . . . . .
A qualitative idea of the magnitude of this gain is seen by comparison of the intensities
of the NiSO4.xH2O peak
at 986cm-1 taken from the best spectra recorded by each
method (figure 5).
| Solid |
Pyrex/mull |
Quartz/mull |
KBr/mull |
KCl disk |
| 0.040 |
0.125 |
0.125 |
1.175 |
0.300 |
These results, used in conjunction with all the other results, show pressed disks are
better at dissipating heat than nujol mulls. Also KCl is better than KBr which in turn is
better than both quartz and Pyrex. This trend, as expected, is directly related to the
thermal conductivities of these media; values of relevant thermal conductivities are 88,
71 and 38 mWcm-1K at 273K for KCl, NaCl and KBr,
respectively when pressed at 8000 atmospheres.
The improvement in the spectra was so pronounced because the effective heat dissipation
provided by contact with a medium of good thermal conductivity allowed higher laser powers
and longer scanning times to be used. To give an example; as a pure solid, KMnO4 decomposes when laser powers greater than 300mW are used.
For all the sampling techniques used, it was possible to record the spectrum using the
maximum available (900mW) laser power. Another illustration is provided by graphite. At
powers of greater than 50mW the spectrum of pure graphite (a very poor scatterer) is
completely dominated by the heating background. When the powder is presented in a KCl
disk, powers of up to 300mW can be employed and the spectrum shows minimal heating
effects; it is also possible to record 1000 scans or more with little change in the
heating background but vast improvement in the quality of the spectrum.
There are a few potential disadvantages of some of these methods. In pressing disks, the
high pressure conditions may result in sample decomposition or degradation as was observed
for CuSO4.5H2O. The
use of quartz and Pyrex flats resulted in an effective low frequency cut-off at 500cm-1 (due to their inability to transmit radiation below this
frequency) and therefore made the use of these less desirable.
There is no equivalent cut-off problem for KBr flats and, furthermore, they are readily
available making them the preferred choice. They do, however, have the disadvantage that
they can fog, depending on the experimental conditions. Spectra of samples run as nujol
mulls are contaminated to some extent by the spectrum of nujol. This was found to be
insignificant in most cases but would be more of a problem if there were bands of interest
in the region 2800-3000cm-1; this could be overcome by
using an alternative mulling agent such as hexachlorobutadiene which has no bands in the
region of interest. No problems were encountered in recording the spectra due the
effective low sample concentration (particularly for the pressed disks) but it should be
borne in mind that the signal intensity does depend on the sample concentration; in some
cases there may need to be a compromise between signal intensity and heating effects to
obtain an acceptable spectrum.
The quality of a Raman spectrum recorded as a nujol mull depends on the particle size,
thickness of the mull between the flats and stodginess of the mull. These factors were
investigated using NiSO4.xH2O
as a nujol mull between quartz flats (thickness) and KBr flats (particle size); each
series of spectra were recorded under identical spectrometer conditions. The results were
analysed in terms of the heating effect (broad background intensity at high frequency) and
peak intensities (comparing the band at 986cm-1).
. . . . . . . . . . . . . . . . . . . . . . . .
[ijvs/volume1/edition1/section2.html#page31] . . . . . . . . . . . . . . . . . . . . . . .
.
Figure 7: Spectra of NiSO4.xH2O
recorded as a nujol mull between quartz flats using 250mW laser power, 4cm-1
resolution and 200 scans to show the effect of mull sample thickness: (a) 0.25mm, (b)
0.62mm, (c) 1.14mm, (d) 1.71mm and (e) 2.33mm.
[High-Resolution Spectrum]
. . . . . . . . . . . . . . . . . . .
. . . . . [ijvs/volume1/edition1/section2.html#page32] . . . . . . . . . . . . . . . . . .
. . . . . .
Figure 7 shows the effect on the spectrum of NiSO4.xH2O due to changing the sample thickness between the flats from
0.25 to 2.33mm. In general, the amount of sample heating increases with increasing sample
thickness. This would be expected because there is poorer contact between the sample and
heat dissipating medium. The amount of heating is also a function of the number of scans
recorded. The spectral intensity appears to be almost independent of the sample thickness,
but very thin samples show slightly weaker signals. Using stodgier mulls marginally
increases the heating effect over that seen for runnier ones (more noticeable for thicker
samples). However, lowering the proportion of nujol has the advantage of giving a
relatively weaker nujol spectrum which necessarily overlays (and therefore contaminates)
the sample spectrum. The effect of particle size in the mull is illustrated in figure 8.
Decreasing the particle size decreases the amount of sample heating and increases the
relative signal intensity. This is consistent with improved contact with the thermal
dissipating medium and the expected increase in the scattering cross-section provided by
the larger available surface area.
Figure 8: Spectra of NiSO4.xH2O
recorded as a nujol mull between KBr flats using 200mW laser power, 4cm-1
resolution and 500 scans to show the effect of particle size: (a) coarse, (b) medium and
(c) fine.
[High-Resolution Spectrum]
Therefore, the optimised mull to minimise heating effects should be a fairly thin
layer between the flats ( 0.5mm or less), quite stodgy (to maximise sample concentration and
minimise nujol effects without compromising heat dissipation) and composed of small
particles.
. . . . . . . . . . . . . . . . . . . . . . . .
[ijvs/volume1/edition1/section2.html#page33] . . . . . . . . . . . . . . . . . . . . . . .
.
Conclusions

It is possible to use the crystal transition of Na2MoO4 at about 112°C to estimate the laser heating effect during
a Raman scan by recording the spectrum of the pure solid. Laser powers of 125mW raised the
sample temperature by 40°C; larger powers were able to raise the temperature by greater
amounts. As sodium molybdate is a fine white powder, the effect is likely to be
significantly greater for coloured samples thus emphasising the need for alternative
sampling techniques for heat sensitive materials. There are also significant temperature
gradients across the sampled region. Thermal equilibration can take in the order of 10
minutes exposure to the laser beam.
The use of nujol mulls and pressed disks can significantly reduce the degree of sample
heating. In general, pressed disks are better than nujol mulls because higher laser powers
can be used and the spectrum is not contaminated with that of nujol but they have the
disadvantages that the sample concentration is relatively low and their preparation could
result in sample degradation or decomposition. For the methods tried, the effectiveness in
minimising sample heating can be qualitatively ranked as:
KCl disk > KBr disk > KBr/mull > quartz/mull Pyrex/mull >> pure solid
The order is consistent with the thermal conductivity of the dissipating medium.
Further work with nujol mulls revealed the amount of sample heating depends directly on
the thickness of the sample layer between the flats and the particle size as well as, to a
lesser extent, the stodginess of the mull. Using thinner samples and smaller particles
reduces the heating effects because of the improved thermal contact with the flats.
Moreover smaller particles give an increased scattering cross-section resulting in
stronger signals. Stodgier mulls showed only slightly higher sample heating effects than
thinner mulls but their spectra were relatively less contaminated by nujol.
To summarise, if solid samples are significantly heated by the laser beam preventing
the successful recording of a Raman spectrum, it is appropriate to try methods normally
associated with infrared spectroscopy. Satisfactory results can generally be obtained
using a liquid paraffin mull between KBr flats; the mull should contain small particles
(well ground sample), be of a fairly stodgy consistency (firmer than toothpaste) and form
a layer <0.5mm thick between the flats. A pressed disk, particularly if made with
spectroscopic grade KCl, would give better results provided the sample does not decompose
in preparation; such disks can be thicker than and contain a greater proportion of sample
than is usual for IR studies.
. . . . . . . . . . . . . . . . . . . . . . . .
[ijvs/volume1/edition1/section2.html#page34] . . . . . . . . . . . . . . . . . . . . . . .
.
Acknowledgements

The author thanks EPSRC for financial support and Prof. P.J. Hendra for many helpful
discussions.
References
- S.J.A. Pope, Y.D. West, Spectrochim. Acta Part A, 51 (1995) 2011.
- P. Hendra, C. Jones, G. Warnes, Fourier Transform Raman Spectroscopy: Instrumentation
and Chemical Applications, p.28, Pub. Ellis Horwood Limited, 1991.
- R.J. Lehnert, Ph.D. Thesis, University of Southampton, 1994.
- S.J.A. Pope, Y.D. West, Spectrochim. Acta Part A, 51 (1995) 2027.
- G.Dent, Spectrochim. Acta Part A, 51 (1995) 1975.
REF: Int.j.Vib.Spect., 1, 1,
5 (1996)