7. The application
of Andy Brookes and Derek Craston, Abstract
Introduction Raman spectroscopy has been applied to an extremely wide variety of chemical systems. The ability to study solids, liquids, gases and solutions non-invasively and non-destructively combined with sampling through glass, optical fibres, microscopes and macro optical arrangements makes Raman an extremely versatile technique. Two excellent reviews cover many of the applications of Raman [1,2]. This paper is specifically concerned with reaction monitoring and aims to provide an indication of some of the uses of Raman in this field. The advantages gained from in situ process monitoring which have been facilitated by the increased use of fibre optic probes and chemometrics has made this an area of increasing importance. IR, NIR and to a lesser extent Raman and ultrasound are all beginning to make an impact as industry moves away from a reactive to a proactive attitude towards analysis. To highlight its versatility Raman was applied to three different
chemical systems. This liquid based (aqueous suspension of styrene) polymerisation reaction has been studied extensively using Raman [3]. Whilst both in situ[4] and flow cell [5] based systems have been used to study polymerisation reactions in the laboratory environment, the data presented here was obtained from a laboratory experimental set-up using a fibre optic probe. The work highlights the industrial applicability of information obtained from Raman experiments and also gives an indication of the large amount of data which can be easily collected when one avoids removing aliquots and working up the sample prior to analysis. Why was Raman chosen to monitor this reaction? The desire for non-invasive, non-destructive real time monitoring limited the choice of analytical tool to spectroscopic techniques. The water present gives rise to problems in the NIR and IR due to its strong spectra but has a very weak Raman spectrum. The vinyl band, the functional group of interest, gives a very strong Raman signal making its disappearance easy to follow. This band is much weaker in the IR spectrum, making its analysis much more difficult (but not impossible). Also in the reaction mixture are additives e.g. initiators, although these are at low concentration there is a possibility that they could act as interferences in some analytical methods. When using Raman their spectra are too weak to interfere with the styrene spectrum. Adhesive Curing Adhesive curing is an area where non-invasive sampling is highly desirable. Fluorescence has been a problem in the analysis of curing systems but the introduction of NIR sources has largely overcome this. Raman has been shown to be capable of monitoring the cure of several adhesive systems such as cyanoacrylate, methacrylate and epoxy [6]. This data has been used to evaluate the kinetics and mechanism of curing as well as evaluating individual components such as crosslinking agents [7]. Composite material degradation As well as the curing of polymeric systems an important issue to manufacturers is degradation [8]. This is particularly important when considering composite materials. These are seen as the materials of the future, combining what are often very different materials to give desirable physical properties [9]. In recent years a wide array of composite materials has been developed. These present a new set of analytical challenges. The extent and cause of degradation of these materials and the effects the degradation has on the chemical and physical properties needs to be investigated thoroughly in order to ensure they are suitable for their proposed applications. The main issue of interest in the analysis discussed here is the ingress of water into the material. Although the exact storage conditions and sample composition is not known it is believed that the samples were a polymer/glass composite and the storage conditions were the same for all of the samples with the only variable being time. Whilst it would seem sensible to study this problem with IR or NIR this was not feasible. As the material degrades it becomes darker in colour. The darkening of the material increases the IR and NIR absorbance of the material at all wavelengths preventing any meaningful IR and NIR analysis. Using visible excitation Raman spectra could not be collected due to fluorescence but FT Raman gave excellent spectra without problems from sample heating. Instrumentation Over recent years the number of instrument manufacturers has increased. Users are no longer restricted to a very limited number of specialist suppliers but can choose from a wide variety of vendors. Most of the new manufacturers have instruments built around diode lasers, CCD cameras and fibre probes. These range in price between $10K and $100K. It is obvious that compromises have to be made when designing an instrument at low cost. These become apparent as limitations in sensitivity, spectral resolution and range. It appears that instrument manufacturers who work in a particular area e.g. document examination or process monitoring, are beginning to recognise the potential of Raman and are designing instruments specific to their application. An extensive list of manufacturers can be found on the Links page of this journal. One point has to be made here: all of the instruments of the type described above are prone to fluorescence problems by their very nature. When these systems are applied to real industrial samples a clearer indication of their usefulness will develop. There are two practical alternatives available for overcoming fluorescence
Experimental Styrene polymerisation Adhesive curing Composite material degradation Results and Discussion Styrene polymerisation
Equation 1 Where Mo is the monomer concentration at time 0 and Mt is the monomer concentration at time t. The area or intensity of the vinyl band can be substituted directly for Mo and Mt without having to calculate the monomer concentration.
The data obtained from each isothermal polymerisation was placed into a kinetic model. The model was then used to optimise the process by calculating the best temperature ramp conditions. A temperature ramp is desirable as these conditions are known to result in the least residual monomer in the polymer. The calculated ramp conditions were also found to give the highest % conversion in the shortest time whilst yielding a polymer with appropriate physical properties e.g. molecular weight distribution. On the basis of this laboratory work a Raman system was installed upon an industrial batch reactor to monitor the production of polystyrene. This is an excellent example of taking an analysis out of the laboratory and placing it into an industrial environment. Adhesive curing
Figure 4 shows the changes in the Cº N region over the course of the reaction. It can be clearly seen that the single Cº N stretch band present, 2235cm-1, decreases in intensity as the reaction proceeds. Associated with this is the appearance and increase in intensity of the band at 2245cm-1. It is likely that this is also due to Cº N. This indicates either that the Cº N bond is being changed chemically, i.e. new substituents are being added to the carbon, or that identical groups are in two very different environments. The increase in intensity of the peak at 2245cm-1 whilst the band at 2235cm-1 is disappearing gives rise to an isosbestic point. This indicates that the two bands are very closely associated and dependent upon each other i.e. there is a one to one conversion from one form or environment to the other.
Figure 5 shows the changes in the region 900-700cm-1. There is a noticeable shift in position of the band at 860cm-1 whilst the band at 840cm-1 decreases in intensity greatly. Although there may be several assignments for the band at 840cm-1, it is possibly due to an oxirane ring (COC).
As well as obtaining information on the chemistry of the reaction it is also possible to obtain information on the extent of cure. Figure 6 shows the data obtained from the decrease in intensity of the 840 cm-1 band. The intensity was expressed as % conversion (equation 1) and plotted against time. Although data for this reaction was only obtained at RT it would have been possible to obtain data at a variety of temperatures, if desired, in order to elucidate kinetic information such as reaction rates and activation energies.
Composite material degradation It is apparent from the spectra that the material has some aromatic content. This is indicated (in the non-degraded spectra 1-2) by, for example, the ring breathing mode of the benzene ring at ~1000cm-1 and the relatively intense aromatic CH stretch band, ~3050-3100cm-1. It is apparent that a large chemical change has occurred between samples 2 and 3. The ring breathing mode is no longer active and the relative intensity of the aromatic CH stretch has decreased. Also apparent is an increase in intensity of the OH stretch region (see expanded view, Figure 8) at ~3250-3500cm-1.
Figure 8 shows that samples 1 and 2 have very similar spectra in the OH stretch (3250-3500cm-1) and CH stretch region (2800-3100cm-1), with very little contribution from OH. The difference between samples 2 and 3 are apparent. The aromatic CH stretch (3050cm-1) to aliphatic CH stretch (2925cm-1) ratio has dropped dramatically. There has also been an increase in the amount of OH (3250-3500cm-1) present in the sample. Sample 4 has a similar aromatic to aliphatic CH stretch ratio but the amount of OH has increased above the level present in sample 3. Associated with these changes are changes in the functional group and fingerprint regions of the spectra. Figure 9 shows these regions (1800-200cm-1). The spectra show the change from the mono, meta or 1,3,5-substituted benzene ring pattern, spectra 1 and 2 (characterised by the intense band at ~1000cm-1), to a pattern which is characteristic of other substitution patterns (characterised by the single intense band at ~1600cm-1), spectra 3 and 4.
This evidence suggests that as the water content of the composite material increases the substitution on the aromatic ring also changes, associated with this is a decrease in the concentration of aromatic CH. This indicates an increase in substitution on the aromatic ring. The exact mechanism for this is unclear i.e. does the ingress of water into the material directly affect the substitution on the ring or does it facilitate another reaction in the material which gives rise to the change in substitution? It is also apparent from the differences between spectra 3 and 4 (Figure 8) that the OH concentration increases but the aromatic to aliphatic CH ratio stays the same, i.e. there is no further substitution on the aromatic ring. This may be due to a secondary reaction taking place which does not affect the aromatic to aliphatic ratio. There are no other changes in the spectra (samples 3 and 4) which are indicative of further chemical reactions. Another explanation is the retention of water by the material. Although the mechanism for both the increased substitution on the aromatic ring and the further increase of the OH content of the material is unclear the effects upon the physical properties of the material are apparent e.g. severe discoloration of the material. Conclusion The styrene polymerisation data presented in this paper highlights the ability of Raman to obtain high quality chemical data from difficult sampling situations i.e. high temperature, high pressure, aqueous solutions. This work also illustrates the increased amount of data which can be obtained when aliquot collection is avoided. This means that continuous monitoring of processes or reactions can be achieved. Any errors associated with the work up procedure when using the aliquot method are also avoided. Also shown is the transferability of laboratory data to an industrial situation. These issues also apply to the study of adhesive curing. It is possible to use Raman to study curing rates in order to establish curing kinetics and hence aid the elucidation of optimum product composition and ideal curing conditions. The use of Raman to investigate degradation proved successful. Although full mechanisms could not be elucidated due to lack of information regarding the composition of the material, it appears that two mechanisms were taking place, one which changed the substitution on the aromatic rings and another which increased the OH content of the material without changing any of the chemical structure. It is hoped that the elucidation of the degradation mechanisms of composite materials could aid in the discovery of alternative formulations which have greater resistance to degradation. Acknowledgements Elements of this paper were supported under contract with the Department of Trade and
Industry as part of the Analytical Innovation Programme. Thanks also go to NPL for
supplying the composite material samples and Actinic Technology for supplying the styrene
polymerisation data.