Following the question and answers between Richard Duerst and
Mike Ford and Andy Turner featured in last issue's Dear Readers [Edition1], there's been more toing and froing
of emails. Richard's response to Mike and Andy's answer came in at
the end of February...
An extension to the first question, i.e. "What are the trade-offs in going from an 8-
to a 16- to a 20-bit A/D convertor?" is: Can the same signal-to-noise ratio be
achieved with a spectrometer with an 8-bit A/D as with a spectrometer with a 16-bit A/D if
signal averaging is employed, with random 'noise' at the level of one or two least
significant bits? I realize that there will be a significant difference in accumulation
time.
Mike then came back with this answer...
I have discussed this question with Dr Andy Turner and we agree the answer. In principle,
it would be possible to achieve the 8 bit improvement in S/N by summing 64,000 scans (the
improvement in S/N goes as the square root of the number of scans added, so you need 2 to
the 16 scans). However, almost invariably, other factors come into play which limit the
maximum improvement to the S:N ratio to much less than 8 bits. Such factors include the
noise not being truly random and non-linearity of the A/D convertor which, ideally, should
be as good as the most significant 8 bits of a 16 bit convertor. Of course, the memory
into which the signal is summed would need to be 24 bits (or the equivalent) in order to
avoid overflow.
Then Richard had another question...
But what are the non-random noise sources relevant to IR?
Mike's reponse this time was...
I have had another chat with Andy and here is the best we can do as a quick answer - we do
not guarantee that it is complete! We have assumed that 'relevant to IR' means FTIR - some
of the things we mention are specific to FT.
Possible non-random 'noise' sources include: anything which causes a repetitive
disturbance of the initial data collection interval - such as unequal mark/space ratio in
the reference laser fringes or a repetitive scan velocity variation; any non-linearity in
the total process of conversion of the optical (IR) signal to its digital form including
non-linearity in any of the components, i.e. the detector, amplifier(s), A/D convertor;
any spurious signal picked up which is synchronous with the scan or data collection
interval. Some of the above might not be truly classified as 'noise', but they will still
cause spectral distortion, as will numerous optical effects.
Back came another question from Richard....
Would the cooling of a detector - it rarely reaches a steady state - and room
temperature variations also be included as non-random noise?
And Mike's answer was.....
The best answer I can give is as follows. Room temperature fluctuations will, almost
certainly, be slow compared with single scan times and their effects (probably minor
changes in the spectral profile) will tend to be averaged out by summing multiple scans.
They are best considered as slow, random noise which will not tend to limit the ability to
improve S:N by multiple scanning. On the other hand, detector temperature
fluctuations can be much faster, due to such things as bubbling of the coolant, and could
be of a similar time scale to scanning. However, it is very unlikely that any such
variations would be even partly synchronised with the scanning, and any associated
noise would therefore be random and would average out.
All seems now quiet from Richard at the moment! Actually it's great that IJVS can get
people conversing with each other (albeit via computer) from all over the world. We're
happy to help.
Quite of few of you have emailed asking about the impact
factor of IJVS and quite right too as we are a bone fide journal.
We have been trying to get an answer out of the American Chemical
Society and in fact wrote an official letter to them at the end of
November 2001, but they haven't replied. So we've tried by email and
still no answer! If you read Patrick's Editorial,
you'll hear his thoughts on the matter! Now on another matter, we
had a plea from Professor Salvatore Romano from Florence, Italy....
I am a professor of medical physics at the University of Florence. I was assigned the job
to develop within one year a laboratory for practical lessons and group operation
beginning with simple to more sophisticated instrument. Say: validation of the Poiseuille
Law to IR-UV-Vis, fluorescence and X and gamma spectrometry. Some of these experiments are
to be confined to a 3-year course for medical-biotechnology students. My personal research
topics are mainly on FTIR and autofluorescence micro-spectrometry and imaging of
biological samples. I ask you for some, even obsolete but still functioning,
instrumentation such as a fluorescence microscope, an FTIR (or grating) spectrometer, a
densitometer (for radiography digital conversion) a beta-counter and so on.
Salvatore Romano, Prof. of Medical Physics, Dept of Clinica Physiopathology
University of Florence, viale G. Pieraccini, 6 50139 Florence - Italy
Phone: +39 (0)55 4271 465 (room), +39 (0)55 4271 /218; /219; /217 (Lab)
Fax: +39(0)55 4271 413
Email: s.romano@dfc.unifi.it
Can anyone help Salvatore? Recently we have advertised the availability of IR kit in
the States. I'm sure some must be lying around in Europe.Please help and keep us informed.
Another request for information came in , this time from Jorge
Villafuerte from Ecuador..
I am interested in information about the analysis Diesel fuel using infrared - the
analysis to be quantitative and qualitative.We want to determine elements such as sulphur,
gum, etc. Can you help me with these problems?????
Again is there anyone who can help Jorg, his email is jorge.j.villafuerte@secuado.shell.com.
If you can, then please don't forget to
copy us here at the Office.
Please keep sending your comments and
questions, sometimes it may take a while for us to answer, but we will.
All your questions and our replies are published here in Dear Readers. |

[IJVS
Home] [Hot
Sources] [Links]
[Bookshelf]
[Contents]
[Publication
Details] [Help] [Back Issues]
[Feedback] [Announcements]

Copyright © 2004 John Wiley & Sons, Ltd
|